138Kr run Aug 2009
Target issues
* Target number #405 UC2C, Ta
VD5+Cu, cold line
* Beam on target (no converter)
* Mass markers: 95% Ar + 5% Xe
* HRS separator
* Extraction electrode 50 mm
* Uht(HRS,CCV)=30.20
kV, Uht(HRS, source)=30.20 kV, Uht(HRS,
HT_target)=30.1997 kV, Uht(trap)=30.00kV,
Elec05=170 V
* Ion source efficiency around 15%
* Yield measurements performed on 138Xe: awaiting input from Giles/Stora
* Count rate at Miniball: awaiting
input from the users
Experimental issues
* Beam to Miniball
* IS415
Tuning
* Slow extraction from EBIS, optimized on 126Xe
* REXTRAP free-running relative to proton impact
ISOLDE
136Xe:
HRS.FC490=6.1nA, HRS.FC748=4.7nA, trap inj
plate=4.37nA
RFQ eff=77%,
HRS-to-REX eff=72%, no slits inserted
Low energy
*
Stable beam 138Xe (normal extraction)
Icoll=190 mA, Tperiod=183 ms, Tbreed=180 ms (almost optimised
for 33+), Tbeamgate=18 ms
Trap
inj plate 3.1 pA, RFQ.FC20(33+)=7.5 epA => total
efficiency=7%
(very
difficult to read as the Keithley is jumping)
* Radioactive beam 138Xe
(slow extraction)
Icoll=185 mA, Tperiod=163 ms, Tbreed=160 ms (too
low, high rep rate)
Trap inj plate
average 42 pA, RFQ.FC20(33+)=58
epA => 4.2%
Linac transmission and energy:
Accelerated
radioactive beams: 138Xe33+(4.1818)
Energy:
2.865MeV/u
+/-0.3% FWHM
RF pulse length 700 us
Transmission:
A/q=4 RFQ.FC20 to XL65.FC50 85%
126Xe30+
A/q=4.2 RFQ.FC20 to XL65.FC50: 73%
Total transmission
HRS.FC490_to_trap x Trap_EBIS x slow_extraction * Linac x
experiment collimator
0.72*0.06*0.95*0.73*0.75=2.5%
Delivered
beams
138Xe33+
15
out of 34 pulses, 8E12 ppp, 0.5 uA
Current
varying significantly
Trap
inj plate 42 pA
RFQ.FC20
average 58 epA
XL65.FC50
average 35 epA (NB. Beam defocused -> not all
recorded)
Setting summary
For HRS we used:
*
HRS-SEP-9-08-18-30340.csv
* RFQ-9-08-18-30200.csv
* RFQ-REX-9-08-18-30340.csv
REX low
energy settings saved as:
*
REX-30kV-2009-08-21_HRS_138Xe.csv
*
BTS-30kV-2009-08-21_HRS_138Xe.csv
*
EBIS_2009-08-21_Aq41818_slowextraction.csv
REX separator and linac
base settings saved as:
*
090820_1730_SEP_Aq4.1818_2.85MeVu_70%_MB.csv
*
090821_1630_Aq4.1818_2.85MeVu_tunedMB.csv
Technical problems
* IHS switched off due to a fuse
for the water cooling tripped. Down time 1 h
* 7gap amplifier stopped. Down
time 15 min
* HRS MAG90 lost its
communication. All units had TGM timeout. Reboot of dsc Disopow sorted out the
situation. Down time 2 h.
* 9GP.MQ30 stopped due to
overheating. Cooling circuit flushed, worked thereafter. Problem occurred
before the actual beam time.
Comments / Lessons learned
* The linac
transmission for heavy and highly charged 126Xe30+,31+,32+ was around 70-75%
while the transmission for light A/q=4 (20Ne5+ and He+) and A/q3.666 (20Ne6+)
was 85-86%. Linac scaling working so not the reason. We suspect that a poor
vacuum, around 1E-6 mbar, in the DB3 / buncher
section causes beam scattering and/or charge exchange. To be verified.
* The users were limited in the amount of beam
they could handle. Thus, a very low proton current on the target of ~0.1 uA was taken. Apparently the users inserted a slit in the linac to reduce the intensity even further.